World Rural Health Conference
Home Print this page Email this page Small font size Default font size Increase font size
Users Online: 2639
Home About us Editorial board Search Ahead of print Current issue Archives Submit article Instructions Subscribe Contacts Login 


 
 Table of Contents 
ORIGINAL ARTICLE
Year : 2019  |  Volume : 8  |  Issue : 11  |  Page : 3718-3725  

Quality of spirometry testing in a community setting: A study among elderly persons in a rural area of Haryana


1 Centre for Community Medicine, India Institute of Medical Sciences, New Delhi, India
2 Department of Pulmonary Medicine and Sleep Disorders, All India Institute of Medical Sciences, New Delhi, India
3 Department of Biostatistics, All India Institute of Medical Sciences, New Delhi, India

Date of Submission30-Aug-2019
Date of Decision17-Sep-2019
Date of Acceptance09-Oct-2019
Date of Web Publication15-Nov-2019

Correspondence Address:
Dr. Ravneet Kaur
Centre for Community Medicine, All India Institute of Medical Sciences, Ansari Nagar, New Delhi - 110 029
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jfmpc.jfmpc_719_19

Rights and Permissions
  Abstract 


Introduction: Chronic obstructive pulmonary disease (COPD) is a leading cause of morbidity and mortality worldwide. Early diagnosis and management of COPD require good quality spirometry testing, which is currently not available at primary care level in India. This study reports the quality of spirometry testing at the community level among elderly persons in a rural area. Materials and Methods: A community-based cross-sectional study was conducted among 449 elderly persons in a rural area of Ballabgarh block of Haryana state by a trained investigator. A portable spirometer (MIR Spirolab®) was used. House-to-house visits were undertaken. A self-developed pretested semistructured interview schedule was administrated and spirometry was done according to the American Thoracic Society/European Respiratory Society (ATS/ERS) guidelines. Results: Acceptable quality of spirometry tests was found among 87.3% (95% CI: 84.2%–90.4%) participants. Poor quality of spirometry was associated with low Body Mass Index (BMI) (aOR = 0.49, 95% CI = 0.26–0.93) and age ≥ 70 years (aOR = 0.45, 95% CI = 0.21–0.94) in multivariable analysis. Conclusion: Acceptable quality of spirometry can be performed in community settings by using a portable spirometer.

Keywords: Community, elderly, quality, rural, spirometry


How to cite this article:
Kaur R, Kumar A, Hadda V, Kalaivani M, Nongkynrih B, Kant S, Gupta SK, Vignesh D. Quality of spirometry testing in a community setting: A study among elderly persons in a rural area of Haryana. J Family Med Prim Care 2019;8:3718-25

How to cite this URL:
Kaur R, Kumar A, Hadda V, Kalaivani M, Nongkynrih B, Kant S, Gupta SK, Vignesh D. Quality of spirometry testing in a community setting: A study among elderly persons in a rural area of Haryana. J Family Med Prim Care [serial online] 2019 [cited 2019 Dec 7];8:3718-25. Available from: http://www.jfmpc.com/text.asp?2019/8/11/3718/270942




  Introduction Top


Chronic respiratory diseases like chronic obstructive pulmonary disease (COPD) and asthma are important health problems in India. These diseases cause 10.9% of total deaths and contribute to 6.4% of all Disability Adjusted Life Years.[1]

As reported by the World Health Organization (WHO), COPD was the third leading cause of death in WHO-SEAR in 2015. Almost 90% of COPD deaths occur in low- and middle-income countries.[2] The magnitude of COPD is substantially higher among elderly persons.[3]

Worldwide, chronic respiratory diseases are under-diagnosed, and about two-thirds of the patients are never diagnosed even in developed countries.[4],[5],[6] Spirometry is an important investigation for early diagnosis and monitoring of patients with chronic respiratory morbidities. It is the most objective and reproducible measure of airflow limitation.[7] A high proportion of COPD in the community remains undiagnosed, as spirometry is under-used, particularly among elderly persons. In India, spirometry is available only at district-level health facilities or above. Access to spirometry at the level of primary care may help in early diagnosis, adequate management, and monitoring the progression of the disease. At the same time, quality of spirometry testing is an important concern. The patient is required to perform a number of rigorous breathing maneuvers, which require motivation and coaching by trained personnel. Therefore, the quality of spirometry tests is critically dependent on the skills of the operator and cooperation of the patient. This may be challenging in community settings, as compared to a laboratory. The results of previous studies conducted worldwide are inconsistent as to whether the quality of spirometry performed at the community level meets adequate minimum standards, especially among elderly persons.[8],[9],[10],[11] In India, there is a paucity of studies on quality of spirometry testing in community settings. The present study was conducted among elderly persons in a rural area of Haryana and aimed to assess the quality of spirometry testing at community level.


  Materials and Methods Top


This community-based, cross-sectional study was conducted in the rural field practice area of a research institute. The field practice area consisted of 28 villages in Ballabgarh block of Faridabad district of Haryana, and had a population of about 1,00,000 individuals, including nearly 7000 elderly persons. There is a computerized database of all individuals residing in the area, which is updated annually.

The study was conducted among elderly persons aged 60 years and above residing in the villages under field practice area for at least past 12 months.

Elderly persons who were seriously ill, unable to comprehend or had any contraindication to perform spirometry (any severe injury or surgery to abdomen, chest, or eye in last 3 months; myocardial infarction in last 3 months; hospitalization due to cardiac illness in last 30 days or were on treatment for tuberculosis) were excluded.

The sample size was calculated using the formula 4pq/d 2. Because studies on the quality of spirometry in India were not available, proportion of persons getting acceptable results was assumed as 50%, and absolute error was taken to be 5%. Considering a nonresponse rate of 15% and migration of 5%, the sample size was increased to 500 elderly persons.

A list of persons aged 60 years and above was taken from the computerized database, and 500 participants were selected by simple random sampling using computer-generated random numbers.


  Details of Spirometery Top


A portable spirometer MIR (Medical International Research) Spirolab ® was used.[12] The spirometer complied with the ATS/ERS criteria for accuracy.[13] It was portable, had a long battery life, incorporated a quality assurance program, had an in-built printer, and it allowed the export of data. A single model of a spirometer was used for all participants.

Training

The investigator (AK) who conducted the spirometry tests was a medical graduate and was trained in the pulmonary laboratory of the research institute for 1 month (150 h) under the supervision of a faculty member of Department of Pulmonary Medicine and Sleep Disorders (VH), who had more than ten years of experience. During the training period, the investigator performed 50 spirometry tests, among patients of all age groups, with the assistance of a pulmonary laboratory technician, and 50 spirometry tests, without any assistance, among elderly patients in the pulmonary laboratory.

Quality assurance

The faculty member in Pulmonary Medicine (VH) was responsible for the training of the investigator and interpretation of results of spirometry tests, while the faculty in Community Medicine (RK) was responsible for supportive supervision in the field during house-to-house visits. Before data collection, the investigator performed the spirometry on elderly patients with the portable MIR Spirolab spirometer, and the findings of which were reconfirmed by the spirometer at the pulmonary laboratory of the research institute. All tests were read by the faculty for interpretation. The spirometer was calibrated at regular intervals.

Patient safety

The spirometry maneuver is generally safe. The primary risk associated with it is fainting among older participants with impaired lung function. To minimize the risk, spirometry was done in a seated position and the investigator was trained to watch for signs of dizziness or other problems and to stop the maneuver if necessary. The risk of infection was minimized by using single-use disposable mouthpieces. The testing was rescheduled to a later date for the participants with obvious upper respiratory infections.


  Acceptability Criteria for Spirometry Test Top


Spirometry testing requires the participant to perform strenuous and precise physical maneuvers. The patient must inhale completely to total lung capacity, and without delay, blast the air from the lungs with maximum effort. The action must be maintained until the lung volume is near residual volume.

Recently published guidelines for spirometry by Joint Indian Chest Society-National College of Chest Physicians (India) have clearly stated within-maneuver acceptability criteria, between-maneuver repeatability criteria, and end of test criteria [14] In our study, these criteria were adhered to while performing spirometry, which was in compliance to ATS/ERS criteria as well.[13]

Thus, the following acceptability criteria were adhered to:

- A good start and satisfactory exhalation (in duration and flow)

- No cough during the first second of exhalation

- No abrupt termination

- Maximal effort provided throughout the maneuver

- No obstructed mouthpiece

- No leak

- No glottis closure that influences the measurement.

Quality of spirometry test as shown by spirometer and its interpretation based on acceptability and repeatability criteria[13],[14]



Operational definitions

The following operational definitions were used in the study:

Chronic cough

Cough on most days for 3 consecutive months or more during the year for the past 2 years.[15]

Chronic phlegm

Phlegm on most days for 3 consecutive months or more during the year for the past 2 years or more.[15]

Dyspnea

Breathlessness while walking, which required the patient to stop or slow down for breathing while walking on the level.[15]

Recurrent wheeze

The occurrence of episodes of wheezing/whistling sounds in breathing associated with breathlessness at least once in a year for past 3 years.[16]

Ever smoker

A person who has smoked at least one cigarette/bidi per day for 1 year or more, or who has smoked hookah/chillum ≥10 times per month for 1 year or more.[17]

Never smoker

A person who has not smoked or who does not qualify in the definition of ever smoker.[17]

Body mass index (BMI)

It was calculated as Weight (in kg)/Arm span (in meter)2 as arm span has been reported to be a more valid indicator of height among the elderly.[18]

BMI was categorized into -

  1. Low (less than 18.5 kg/m 2)
  2. Normal (18.5–24.9 kg/m 2)
  3. High (25 kg/m 2 and above).


Acceptable quality test

Test quality A, B, C, and D were considered acceptable.

Unacceptable quality test

Test quality E and F were considered unacceptable.


  Data Collection Top


House-to-house visits were undertaken. In case a participant was not found at home despite three visits, s/he was considered as a nonrespondent. All participants were provided with a participant information sheet, informed about the purpose of the visit, and written informed consent was taken. A pretested semistructured interview schedule was administered in Hindi; information was collected regarding sociodemographic details, history of chronic respiratory symptoms or respiratory disease, and smoking. Weight, height, and arm span were measured using standard techniques.[19] Among persons who were on regular treatment with bronchodilators, spirometry was done on the next day after asking them not to use bronchodilators for 24 h. For smokers, the spirometry test was done after 1 h of smoking free interval.

Spirometry was performed for the assessment of airway obstruction. The measurements were done using a hand-held portable spirometer (MIR Spirolab) according to the standard guidelines (American Thoracic Society/European Respiratory Society)[13] and Joint Indian Chest Society-National College of Chest Physicians (India) guidelines for spirometry.[14]

For each participant, information on weight and height was entered in the spirometer. The participant was asked to sit comfortably. Two measurements—one each of pre- and post-bronchodilator—were performed 20 min apart. Four puffs (100 mcg each puff) of salbutamol were administered via metered-dose inhaler with a double valve Volumatic spacer with a participant in seating position. Separate disposable mouthpieces were used for each participant. Prebronchodilator spirometry was performed in which the participants were asked to repeat the test until Quality “A” results were obtained or the participant was exhausted, whichever was earlier. After this, the participants were asked to inhale 4 puffs of 100 mcg each salbutamol through a metered-dose inhaler using a spacer device. Bronchodilator reversibility was undertaken on all patients, even if prebronchodilator lung function was within the normal range. Postbronchodilator spirometry was performed after a gap of 20 min of salbutamol inhalation. Again, participants were asked to repeat the test until Quality “A” results were obtained or the participant exhausted, whichever was earlier.


  Ethical Considerations Top


The study was conducted after obtaining ethical approval from the Ethics Committee of the research institute (Approval no. IEC-168 dated 25.08.2019). Participants who were found to have airway obstruction were referred to an appropriate health facility.

Statistical analysis

Data were entered in MS Excel 2013 and statistical analysis was carried out using Stata version 11 (College Station, Texas, USA). Sociodemographic profile and clinical profile were reported as mean (standard deviation) or number (percentage). Quality of spirometry was reported as number and percentage in each quality category. Quality categories A–D were included in acceptable quality, while quality category E and F were considered unacceptable. Respiratory symptoms were grouped and analyzed as persons with respiratory symptoms and persons without respiratory symptoms.

A multivariable logistic regression model was used to look for association between spirometry quality and other sociodemographic and clinical factors. First, bivariable logistic regression analysis was performed. Variables with P value < 0.25 were included in multivariable logistic regression model.


  Results Top


Of total 500 randomly selected participants, one had died, three had migrated, while one participant was excluded due to rib fracture. Out of 495 eligible elderly, 44 participants were not available despite three visits, and two participants refused to give consent. Thus, the response rate was 90.7%.

The mean age of the 449 elderly participants was 68.6 years (SD = 6.8 years). Women constituted 57% of the study participants. Majority of the participants (75.5%) were illiterate, 63% belonged to lower-middle socioeconomic status, and 82.9% were married. Majority of the participants (93.3%) were economically dependent on care-providers. Nearly half (50.3%) of the participants were ever smoker. Nine persons (2% of participants) had a history of Antitubercular Treatment (ATT). Dyspnea was reported by 25.4% of the elderly, while 13.6% and 11.4%% of them reported chronic cough and chronic phlegm, respectively. For analysis, we grouped the persons with selected self-reported respiratory symptoms, and analyzed as respiratory symptoms present (n = 128, 28.5%) or absent (n = 321, 71.5%) [Table 1].
Table 1: Characteristics of study participants

Click here to view


In our study, 87.3% of the participants were able to produce acceptable spirometry curve data. The majority were of spirometry quality A (n = 322, 71.7%). Remaining 57 (12.7%) participants could not produce acceptable quality of spirometry curve [Table 2].
Table 2: Distribution of elderly persons by quality of spirometry test

Click here to view


In bivariable analysis, age, literacy, and BMI were significantly associated with quality of spirometry. On multivariable analysis, age more than or equal to 70 years and low BMI showed significant association. Odds of getting an acceptable quality of spirometry was 55% less likely among persons aged 70 years (aOR: 0.45, 95% CI: 0.21–0.94) and above compared to persons aged 60–64 years, and 50% less likely in elderly persons with low BMI (aOR: 0.49, 95% CI: 0.26–0.93) compared to normal BMI [Table 3].
Table 3: Factors associated with acceptable quality spirometry test

Click here to view



  Discussion Top


This study was undertaken in a rural setting among elderly persons. Pulmonary function tests are the most important investigation in the case of chronic respiratory disorders. Ability to perform spirometry in primary care settings benefits both patients, as they do not have to travel to the district hospital, and the health system, as the burden is reduced at the district level. However, obtaining good quality spirometry tests may be challenging in community settings. In our study, acceptable quality of spirometry tests was found among 87.3% (95% CI: 84.2%– 90.4%) participants. Unacceptable spirometry was associated with low BMI (aOR = 0.49, 95% CI = 0.26–0.93) and age ≥ 70 years (aOR = 0.45, 95% CI = 0.21–0.94). This finding is similar to some other community-based studies. Querioz et al. (2018) found acceptable results in 75.9% participants in a community-based study conducted among 255 elderly persons in Brazil. It was reported that older age and lower MMSE (Mini Mental State Examination) score were significantly associated with poor quality spirometry.[20] In a study by Belo et al. among 307 persons in elderly care centers in Lisbon in 2018, acceptable quality spirometry was present in 85.3%. The study followed the ATS criteria and spirometry was done by qualified and certified technicians.[21] Haynes assessed the ability to perform spirometry among elderly persons in a hospital in New Hampshire, England. The study reported that 94.6% of the elderly performed acceptable spirometry by ATS/ERS criteria.[22] This proportion is higher as compared to our study, which might be due to the fact that it was a retrospective review of hospital-based pulmonary function test (PFT) laboratory data. The study also reported that the quality of spirometry test results achieved by elderly persons was comparable to younger adults. Bellia et al. conducted a multicentric case-control study among elderly persons in twenty-four pulmonary or geriatric institutions, distributed throughout Italy, to assess the quality of spirometry among those with previously diagnosed respiratory morbidity (case), and those without previously diagnosed respiratory morbidity (control). Spirometry was done by trained staff. Acceptable spirometer readings were obtained in 83.6% cases and 81.9% controls. Stepwise logistic regression showed that lower performance was associated with lower activities of daily living score, lower MMSE score, and worse performance in the 6-min walk test. However, older age was associated with poorer FEV1 and FVC, but not poor spirometry performance.[23]

Older age and low BMI were associated with unacceptable spirometry results among the participants.[24] This finding is similar to our study, which suggests that spirometry might be challenging in frail elderly persons. The study also suggested that unacceptable curves might still be clinically useful, as FEV1/FEV6 was shown to have reproducible results and good correlation with FEV1/FVC among elderly persons.[24]

In a study by Turkeshi et al. among two primary care cohorts of community-dwelling elderly in Russia and Belgium, spirometry tests of acceptable quality as per ATS/ERS criteria could be achieved by 43.3% and 57.7% of the elderly persons, respectively. The low percentage of acceptable quality tests may be because elderly of age 80 years or more were studied. Female sex, lower education, depression, and lower MMSE score were reported to be associated with poor quality spirometry. However, the study concluded that impaired cognition measured by MMSE may not be an independent predictor of poor-quality spirometry, even among those who were over 80 years old, and recommended that spirometry should be used more often among the elderly in primary care settings.[25]

Earlier research has highlighted that spirometry in primary care setting avoids waiting time for testing at hospital laboratories, is more convenient to the patients, and also provides timely data to physicians. It can be a viable option at primary health care level provided that it operates under appropriate quality assurance guidelines. The position statement developed by the Canadian Thoracic Society (2013) provided guidance on key factors affecting the quality of spirometry testing in primary care settings. It was recommended that the equipment used should be standardized and the person conducting the test should be appropriately trained. Moreover, interpretation of the tests should be done by an expert.[26]

In their study on the establishment of community-based spirometry service in the Canterbury region of New Zealand, Epton et al. showed that laboratory-quality spirometry can be carried out in the community. They also highlighted the need for adequate training and monitoring of quality assurance to achieve acceptable quality spirometry testing.[27]

In a study conducted among persons at risk of developing COPD in primary care settings of Tasmania, it was reported that early diagnosis of COPD increased with the provision of spirometry testing, performed according to the standard guidelines for quality assurance. The study also concluded that assistance in the form of interpretation of tests by experts is required in such settings.[28]

Primary care settings are usually the first point of contact for patients, particularly in rural areas. Respiratory symptoms such as cough, wheezing, shortness of breath as well as conditions like follow-up of asthma and COPD are among the top-20 reasons for which patients consult a doctor in primary care practice in India.[29] Symptom-based diagnosis of COPD in primary care may be unreliable and leads to inappropriate management of the disease.[30]

It is been reported that a high proportion of patients of COPD are not diagnosed correctly until they reach tertiary care health facilities. A study conducted in Kerala reported that out of 129 patients of COPD diagnosed at a tertiary care hospital, only one had been diagnosed earlier.[31] Hence, good quality spirometry is essential at the level of primary health care, in order to make an early and accurate diagnosis of COPD, which will lead to appropriate management of disease at an early stage.

A recently published report of the Global Initiative for Chronic Obstructive Lung Disease (GOLD 2019) has suggested that good quality spirometry measurement is possible in any health care setting and all professionals who treat patients of COPD should have access to good quality spirometry.[7]

In our study, spirometry testing was done as per the standards for acceptability and repeatability. The investigator received appropriate training at the PFT laboratory, and interpretation of the test results was done by an expert.

However, our study had few limitations. Cognitive impairment was not measured by MMSE scores. Hence, the association of poor cognition with quality of spirometry test could not be assessed. In our study, persons who were not able to comprehend the questions during the interview were proposed to be excluded from the study. However, we did not come across any such participant.

Spirometry testing was done by a medical graduate who was trained to conduct spirometry and was supportively supervised during training as well as during house-to-house visits. The same may not be feasible in other settings.

High response rate and good quality control were major strengths of our study. The investigator was duly trained and the instrument was standardized.


  Conclusion Top


Acceptable quality of spirometry can be achieved among the elderly at the level of primary care by using a portable spirometer. However, it requires comprehensive training and monitoring of quality assurance. The costs of training, equipment, and quality assurance may prove to be potential barriers for providing spirometry testing services at primary care settings.

Declaration of patient consent

The authors certify that they have obtained all appropriate patient consent forms. In the form, the patients have given their consent for their images and other clinical information to be reported in the journal. The patients understand that their names and initials will not be published and due efforts will be made to conceal their identity, but anonymity cannot be guaranteed.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Salvi S, Kumar GA, Dhaliwal RS, Paulson K, Agrawal A, Koul PA, et al. The burden of chronic respiratory diseases and their heterogeneity across the states of India: The Global Burden of Disease Study 1990-2016. Lancet Glob Health 2018;6:e1363-74.  Back to cited text no. 1
    
2.
Global Health Estimates 2015: Deaths by Cause, Age, Sex, by Country and by Region, 2000-2015. Geneva: World Health Organization; 2016.  Back to cited text no. 2
    
3.
Lowery EM, Brubaker AL, Kuhlmann E, Kovacs EJ. The aging lung. Clin Interv Aging 2013;8:1489-96.  Back to cited text no. 3
    
4.
Hill K, Goldstein RS, Guyatt GH, Blouin M, Tan WC, Davis LL, et al. Prevalence and underdiagnosis of chronic obstructive pulmonary disease among patients at risk in primary care. Can Med Assoc J2010;182:673-78.  Back to cited text no. 4
    
5.
Bednarek M, Maciejewski J, Wozniak M, Kuca P, Zielinski J. Prevalence, severity and underdiagnosis of COPD in the primary care setting. Thorax 2008;63:402-7.  Back to cited text no. 5
    
6.
Balcells E, Antó JM, Gea J, Gómez FP, Rodríguez E, Marin A, et al. Characteristics of patients admitted for the first time for COPD exacerbation. Respir Med 2009;103:1293-302.  Back to cited text no. 6
    
7.
Global Initiative for Chronic Obstructive Lung Disease, Incl. Global strategy for the diagnosis, management, and prevention of Chronic Obstructive Pulmonary Disease. 2019 Report. Available from: www.gold.org. [Last accessed on 2019 Sep 20].  Back to cited text no. 7
    
8.
Yawn BP, Enright PL, Lemanske RF Jr, Israel E, Pace W, Wollan P, et al. Spirometry can be done in family physicians' offices and alters clinical decisions in management of asthma and COPD. Chest 2007;132:1162-8.  Back to cited text no. 8
    
9.
Enright P, Schermer T. Don't pay for poor quality spirometry tests. Prim Care Respir J 2013;22:15-6.  Back to cited text no. 9
    
10.
Eaton T, Withy S, Garrett JE, Mercer J, Whitlock RM, Rea HH. Spirometry in primary care practice: The importance of quality assurance and the impact of spirometry workshops. Chest 1999;116:416-23.  Back to cited text no. 10
    
11.
Tuomisto L, Jarvinen V, Laitinen J, Erhola M, Kaila M, Brander P. Asthma programme in Finland: The quality of primary care spirometry is good. Prim Care Respir J 2008;17:226-31.  Back to cited text no. 11
    
12.
Medical International Research. New Spirolab ®. Available from: https://www.spirometry.com/Eng/Products/spirolabNew.asp.[Last accessed on 2019 May 22].  Back to cited text no. 12
    
13.
Culver BH, Graham BL, Coates AL, Wanger J, Berry CE, Clarke PK, et al. Recommendations for a standardized pulmonary function report-An official american thoracic society technical statement. Am J Respir Crit Care 2017;11:1463-72.  Back to cited text no. 13
    
14.
Aggarwal AN, Agarwal R, Dhooria S, Prasad KT, Sehgal IS, Muthu V, et al. Joint Indian Chest Society-National College of Chest Physicians (India) guidelines for spirometry. Lung India 2019;36:S1-35.  Back to cited text no. 14
    
15.
Chhabra P, Sharma G, Kannan AT. Prevalence of respiratory disease and associated factors in an urban area of Delhi. Indian J Community Med 2008;33:229-32.  Back to cited text no. 15
[PUBMED]  [Full text]  
16.
Ly NP, Gold DR, Weiss ST, Celedón JC. Recurrent wheeze in early childhood and asthma among children at risk for atopy. Pediatrics 2006;117:e1132-8.  Back to cited text no. 16
    
17.
World Health Organization. Guidelines for Controlling and Monitoring the Tobacco Epidemic: World Health Organization; 1998. Available from http://www.thl.fi/publications/ehrm/product1/section7.htm. [Last accessed on 2019 May 15].  Back to cited text no. 17
    
18.
Rabe B, Thamrin MH, Gross R, Solomons NW, Schultink W. Body mass index of the elderly derived from height and from armspan. Asia Pac J Clin Nutr 1996;5:79-83.  Back to cited text no. 18
    
19.
World Health Organization. Guide to Physical Measurements. Available from: https://www.who.int/ncds/surveillance/steps/Section%204%20Step%202%20Physical%20Measurements.pdf.[Last accessed on 2019 July 31].  Back to cited text no. 19
    
20.
Queiroz RS de, Faria LM de A, Carneiro JAO, Coqueiro R da S, Fernandes MH. Age and mini-mental state examination score can predict poor-quality spirometry in the elderly: A cross-sectional study. Clinics 2018;73:e374.  Back to cited text no. 20
    
21.
Belo J, Palmeiro T, Caires I, Papoila AL, Alves M, Carreiro-Martins P, et al. Reference values for spirometry in elderly individuals: A cross-sectional study of different reference equations. Multidiscip Respir Med 2018;13:4.  Back to cited text no. 21
    
22.
Haynes JM. Pulmonary function test quality in the elderly: A comparison with younger adults. Respir Care 2014;59:16-21.  Back to cited text no. 22
    
23.
Bellia V, Pistelli R, Catalano F, Antonelli-Incalzi R, Grassi V, Melillo G, et al. Quality control of spirometry in the elderly. The SA.R.A. study. SAlute Respiration nell'Anziano. Respiratory health in the elderly. Am J Respir Crit Care Med 2000;161:1094-100.  Back to cited text no. 23
    
24.
Bhatt SP, Kim Y, Wells JM, Bailey WC, Ramsdell JW, Foreman MG, et al. FEV1/FEV6 to diagnose airflow obstruction-Comparisons with computed tomography and morbidity indices. Ann Am Thorac Soc 2014;11:335-41.  Back to cited text no. 24
    
25.
Turkeshi E, Zelenukha D, Vaes B, Andreeva E, Frolova E, Degryse JM. Predictors of poor-quality spirometry in two cohorts of older adults in Russia and Belgium: A cross-sectional study. NPJ Prim Care Respir Med 2015;25:15048.  Back to cited text no. 25
    
26.
Coates AL, Graham BL, McFadden RG, McParland C, Moosa D, Provencher S, et al. Spirometry in primary care. Can Respir J2013;20:13-21.  Back to cited text no. 26
    
27.
Epton MJ, Stanton JD, McGeoch GRB, Brett I, Shand BI, Maureen P, et al. The development of a community-based spirometry service in the Canterbury region of New Zealand: Observations on new service delivery. NP J Prim Care Respir Med 2015;25:15003.  Back to cited text no. 27
    
28.
Walters JA, Hansen EC, Johns DP, Blizzard EL, Walters EH, Wood-Baker R. A mixed methods study to compare models of spirometry delivery in primary care for patients at risk of COPD. Thorax 2008;63:408-14.  Back to cited text no. 28
    
29.
Rahman SMF, Angeline RP, Cynthia S, David K, Christopher P, Sankarapandian V, et al. International classification of primary care: An Indian experience. J Family Med Prim Care 2014;3:362-7.  Back to cited text no. 29
[PUBMED]  [Full text]  
30.
Vanjare N, Chhowala S, Madas S, Kodgule R, Gogtay J, Salvi S. Use of spirometry among chest physicians and primary care physicians in India. NPJ Primary Care Respiratory Medicine 2016;26:16036.  Back to cited text no. 30
    
31.
Padmanabhan A, Nair S, Jilisha G, Anand J, Babu V, Moosan H, et al. Assessing health-seeking behaviour among asthma and COPD patients in urban South India. J Family Med Prim Care 2019;8:2714-9.  Back to cited text no. 31
    



 
 
    Tables

  [Table 1], [Table 2], [Table 3]



 

Top
   
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
   Abstract
  Introduction
   Materials and Me...
   Details of Spiro...
   Acceptability Cr...
  Data Collection
   Ethical Consider...
  Results
  Discussion
  Conclusion
   References
   Article Tables

 Article Access Statistics
    Viewed101    
    Printed0    
    Emailed0    