ORIGINAL ARTICLE
Year : 2020  |  Volume : 9  |  Issue : 12  |  Page : 6135-6139  

Dosimetric comparison of left sided whole breast irradiation with Tangential wedge beam,electron boosted Tangential wedged beam and asymmetric technique


1 Department of Radiation Oncology, Shohadaye 7 Tir Hospital, Iran University of Medical Sciences, Tehran, Iran
2 Radiation Oncology Research Center, Cancer Institute, Imam Khomeini Hospital; Department of Radiation Oncology, Cancer Institute, Imam Khomeini Hospital Complex, Tehran University of Medical Sciences, Tehran, Iran

Date of Submission18-Jul-2020
Date of Decision15-Sep-2020
Date of Acceptance03-Nov-2020
Date of Web Publication31-Dec-2020

Correspondence Address:
Dr. Mahboobeh Alamolhoda
Department of Radiation Oncology, Shohadaye 7 Tir Hospital, Iran University of Medical Sciences, Tehran
Iran
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/jfmpc.jfmpc_1476_20

Rights and Permissions
  Abstract 


Background: Irradiation of the adjacent critical structures is inevitable in breast cancer radiotherapy (RT). Our purpose is to assess the dose distribution across the breast tissue and adjacent organs with our institutional asymmetric technique for left-sided breast cancer compared to the standard tangential wedged beam (TWB) and electron-boosted TWB techniques. Materials and Methods: The three RT planning were created for 30 consecutive patients with a focus on proper coverage of the planning target volume (PTV). The irritated doses into the heart, ipsilateral lung, and left anterior descending artery (LAD) were evaluated. Results: No significant difference was found in the mean values of relative PTV irradiated to 47.5 Gy, PTV dose and the volume of PTV, and critical organs between the treatments. The mean dose (Dmean) irradiated to the heart and LAD was lowest with the electron-boosted TWB. The Dmean to the heart was comparable between the TWB and asymmetric RT techniques, while the Dmean to LAD was significantly reduced with asymmetric technique versus TWB. The heart volume receiving ≥25 Gy and the Dmean to the left lung were significantly decreased with the asymmetric technique compared with TWB. The mean relative lung volume irradiated to ≥20 Gy was comparable between all techniques. The mean central lung distance was also significantly increased from 18.03 ± 4.5 cm with asymmetric RT to 37.47 ± 5.6 cm with TWB and to 27.67 ± 3.8 cm with electron-boosted TWB techniques. Conclusion: The asymmetric technique is useful for patients with breast cancer on the left side, having acceptable PTV coverage and considerably reduced cardiopulmonary doses.

Keywords: Asymmetric therapy, breast cancer, tangential wedged beam


How to cite this article:
Fadavi P, Ahmadi HK, Diba AA, Jafari F, Alamolhoda M. Dosimetric comparison of left sided whole breast irradiation with Tangential wedge beam,electron boosted Tangential wedged beam and asymmetric technique. J Family Med Prim Care 2020;9:6135-9

How to cite this URL:
Fadavi P, Ahmadi HK, Diba AA, Jafari F, Alamolhoda M. Dosimetric comparison of left sided whole breast irradiation with Tangential wedge beam,electron boosted Tangential wedged beam and asymmetric technique. J Family Med Prim Care [serial online] 2020 [cited 2021 Jan 25];9:6135-9. Available from: https://www.jfmpc.com/text.asp?2020/9/12/6135/305595




  Introduction Top


Surgery can be considered to treat breast cancer or any disorders in the breast or at distant sites in early stages, however, the disease deposits may remain undetected, leading to life-threatening recurrence if remain untreated. Recent studies have shown that radiotherapy (RT) for breast cancer in early stages can reduce the recurrence and mortality rates.[1] However, follow-up studies have shown that RT for left-sided breast has long-term side effects and can increase the risk of ischemic heart disease and pulmonary complications due to the inevitable heart and lung irradiation.[2],[3],[4]

Hence, new therapeutic techniques have been developed to obtain a uniform dose distribution in treatment volume and lower doses to adjacent organs at risk (OARs).[5],[6] One of the treatment methods is tangential wedged beam (TWB) technique, in which two tangential photon beams are used to reduce the risk of irradiation to the heart and lung. To improve target dose homogeneity, wedges are often used.[7]

Modulated electron radiotherapy (MERT) has recently been applied to treat superficial targets including superficial tumors.[8] Electron beams with the ability to modulate the intensity or energy/intensity are available to confirm the target dose.[9]

Despite obvious advantages of intensity-modulated electrons/photons over TWB treatment,[9] more research should be done toward beam planning and delivery for electrons and photons. Using a combination of modulated electrons/photons is the ideal scenario, for example, the use of modulated electron radiotherapy (MERT) for tumor bed boosts and photons for the whole breast irradiation. MERT in combined with intensity-modulated radiation therapy (IMRT) technique is shown to have advantages over the TWB technique and can reduce the irradiated dose to the heart and the ipsilateral lung.[9]

RT techniques vary between institutions depending on their facilities. Considering the nonavailability of multileaf collimators, we designed planning using two usual tangential fields. In addition, medial and lateral beams were copied, but for decreasing the heart irradiation, the width of each beam was decreased by the asymmetric jaws as the asymmetric field.

Here, we performed a comparative dosimetric study to assess our institutional asymmetric method, electron-boosted TWB, and TWB in the treatment of left-sided breast cancer. We analyze the treatment plans with a focus on dose decrease and target dose homogeneity in the heart and lung.


  Materials and Methods Top


Patients and computed tomography scanning

Thirty consecutive patients with cancer in their left-sided breast, who were sent to conduct adjuvant RT after breast-conserving surgery at the Shohadaye Haft-e Tir Hospital, were included between 2018 and 2019. The participants signed the informed consent. The study was approved by the Iran University of Medical Sciences.

To treat the patients, they posited in a supine position while their head turned slightly into the contralateral side and their arms elevated above the head. All the patients with left-sided breast cancer were selected to examine the influence of radiation therapy on the heart. In order to minimize the daily set-up errors and reproducing the daily situations, a custom-made immobilization device such as wing, vacuum, or T-board was utilized. Computed tomography (CT) simulation images with 5-mm thickness were obtained from the neck to the middle of the abdomen to insure the whole radiation area is included. Virtual reality-type techniques were used to design the treatment fields based on the CT-simulation data set. In this study, photon beams with 6 MV and 15 MV were used and obtained from an accelerator (Siemens Primus). To consider the respiratory motion effect in the photon treatment plans, up to 1 cm margin in the posterior direction was added to the planning target volume (PTV).

Treatment planning details

  1. TWBs
  2. In the conventional tangential photon treatment, two physical wedges were used for two beams with aligned bottom edges and gantry angles of 300° and 235°
  3. Two tangential photon beams were combined with an electron beam with enough energy. The electron beam was adjusted until an angle of 5°–10° was obtained laterally from the medial-lateral photon beams. The gantry angles of the two tangential photon beams were larger in comparison with its sizes in the conventional tangential fields. In addition, the length of electron and photon beams was equal. The electron field was 10 or 12 Mev, respectively. Medial electron field is also another common technique used to cover the IM nodes which is well tailored to shallow photon tangents
  4. In our new planning technique, patients were treated with four beams: two tangential beams are used as the same as usual breast planning technique. In addition, medial and lateral beams were copied, and for sparing the heart from excessive radiation, the width of each copied beam was decreased as an asymmetric field. For uniform target dose distribution, we used 6-MV or 15-MV photon beams and 15° or 30° wedges. The angle of gantry for tangential asymmetric beams (medial) was more than of symmetric beams, but the angle of gantry for the other asymmetric beams (lateral) was less than symmetric beams. In addition, beam weighting in this technique in comparison to conventional technique was different.


For treating the PTV, specific objectives were established to treat 90% of the PTV with an ideal dose of 47.5 Gy but the minimum of 45 Gy and maximum hotspots of ≤107%. PTV should receive a mean dose (Dmean) within 100% of the prescribed dose and should not exceed 101.5%. The irradiated dose to OAR was held as low as possible so that the coverage criteria of PTV are not violated, even in the case of exceeding the dose limit of OAR. Treatment planning of radiation therapy was performed using the Isogray (version 4.1.3.23 L) treatment planning system. A dose of 50 Gy in 2.0 Gy per fraction should be delivered to the breast five times a week. Aiming to reduce the interobserver variability, one radiation oncologist did all the contouring. One medical physicist also designed the treatment plans.

Statistical analysis

The irradiated doses to PTV and endangered organs were evaluated. The dose–volume histograms (DVHs) were used to obtain the mean and maximum doses as well as the volume size. The measurement of central lung distance (CLD) was done in a beam's eye view. Analysis of the differences was done in SPSS Statistics for Windows, version 21.0 (SPSS Inc., Chicago, Ill., USA) software using paired t-test. P < 0.05 was considered to be statistically significant.


  Results Top


The dosimetric plans for left-sided breast cancer were studied for the 30 patients using three different techniques.

Volumes

All delineated volumes are shown in [Table 1]. There was not any significant difference in the mean volume of PTV between the three RT methods (708.19 ± 307.5 in TWB vs. 708.17 ± 307.5 in electron-boosted TWB vs. 685.27 ± 325.4 in asymmetric method, P > 0.05). In addition, for the delineated volumes of the heart, left anterior descending artery (LAD), and left lung, no significant difference was found.
Table 1: The mean±standard deviation volume (cm3) for target volumes and endangered organs

Click here to view


Treatment planning data for planning target volume

For all the three techniques of TWB (102 ± 1.44), electron-boosted TWB (103.57 ± 3.67), and asymmetric (101.6 ± 1.8), the mean PTV dose was comparable and close to 100% (P > 0.05). The mean respective PTV irradiated (V47.5) was 93.9%+-3.4% in TWB, 91.97%+-8.83%, and 92.43%+-5.1% in asymmetric RT plan (P > 0.05). Moreover, no significant difference was found in the V50 among all the three techniques. The V45 for asymmetric method was significantly less than the TWB method (97.18% ± 1.84% vs. 98.41% ± 1.34%, P = 0.001), while there was not any significant difference between the V45 of asymmetric and electron-boosted techniques (97.18% ± 1.84% vs. 97.50% ± 2.63%, P = 0.51) [Table 2]. [Figure 1],[Figure 2],[Figure 3],[Figure 4] show the corresponding DVHs. As shown in these figures, there is a tangible reduction in the dose irradiated to the heart, the ipsilateral lung, and the LAD artery.
Figure 1: Histograms indicating the dose versus volume for the heart

Click here to view
Figure 2: Histograms indicating the dose versus volume for the left anterior descending artery

Click here to view
Figure 3: Histograms indicating the dose versus volume for the ipsilateral lung

Click here to view
Figure 4: Histograms indicating the dose versus volume for the planning target volume

Click here to view
Table 2: Therapy planning data for the planning target volume and endangered organs

Click here to view


Cardiac doses

The mean irradiation dose to the LAD (37.04 ± 21.9) and heart (11 ± 5.5) was lowest in the electron-boosted TWB. The average dose to the heart was comparable between the TWB and asymmetric RT techniques (15.65 ± 5.43 vs. 23.2022.54, P = 0.07), while the Dmean of LAD was significantly reduced in asymmetric method compared with TWB (55.21 ± 18.7 vs. 60.55 ± 16.05, P = 0.03).

The mean relative volume of heart irradiated to 25 Gy was significantly decreased with asymmetric technique compared to the TWB (12.93% ± 5.7% vs. 14.82% ± 6.3%, P = 0.004). However, it was significantly larger for asymmetric technique in comparison to the electron-boosted TWB (12.93% ± 5.7% vs. 6.73% ± 5.8%, P < 0.0001). The mean irradiation dose to relative LAD volume (≥25 Gy) was comparable between all the three techniques (57.33% ± 18.77% in TWB, 45.9% ± 90.6% in electron-boosted TWB, and 57.14% ± 19.43% in asymmetric, P > 0.05) [Table 2].

Pulmonary doses

The Dmean irradiated to the left lung was less in the asymmetric (25.85 ± 5.97) and electron-boosted TWB (25.53 ± 7.6) techniques compared with the TWB technique (27.88 ± 5.92) (P = 0.001 and P = 0.01, respectively). However, no significant difference was found between the asymmetric and electron-boosted TWB techniques in terms of average dose to the ipsilateral lung (P = 0.75).

The mean relative lung volume irradiated to ≥20 Gy was 26.28% ± 6.5% for asymmetric method, which was comparable to that for TWB (27.32% ± 6.3%, P = 0.1) and electron-boosted TWB (24.4% ± 8.3%, P = 0.09) techniques. The mean CLD was significantly increased from 18.03 ± 4.5 cm with asymmetric RT to 37.47 ± 5.6 cm with TWB and to 27.67 ± 3.8 cm with electron-boosted TWB techniques (P < 0.0001 for both). The mean CLD was also significantly increased with TWB technique compared with the electron-boosted TWB technique (P < 0.0001) [Table 2].


  Discussion Top


RT treatment for left breast cancer affects the adjacent critical structures, especially heart and lungs.[2],[3],[4] In addition, the breast tissue nonhomogeneous contour results in nonhomogeneity dose distribution and hotspots. The best treatment design would be to achieve a more homologous dose distribution across the breast tissue and a minimum dose irradiated to the OARs.[10],[11] RT techniques vary between institutions depending on their facilities. Here, we compared our institutional RT plan with two other left-sided breast RT techniques (tangential breast irradiation with and without electron boost) in terms of the cardiac and pulmonary dose reduction and dose distribution homogeneity of the target organ.

The tangential breast irradiation is a TWB technique in which the irradiation of the heart and ipsilateral lung tissue is usually inevitable. The field-in-field technique, also known as forward IMRT, is one of the best techniques for sparing OAR with comparable PTV dose coverage levels.[12],[13] However, it cannot be performed in centers without a multileaf collimator. Combined photon and electron beam therapy is an alternative technique with the ability to reduce the OAR radiation.

In this method, trying to separate the heart and lung tissue from central and lateral tangential parts leads to an increase of the gantry angle while the dose reaching the medial portion of the breast tissue is decreased. The electron field is used to compensate this dose deficit because of the low density of this part of the breast. However, optimizing the dose distribution is challenging in this method and daily match line control is mandatory.[14],[15]

Our institutional RT technique is the combination of asymmetric and symmetric fields with common that also spares the heart and lung. Moreover, there is no need for frequent control of match lines or special facilities such as multileaf collimator. Although the electron-boosted TWB technique had the lowest Dmean of LAD (37.04 ± 21.88), our asymmetric RT plan had a significantly lower mean LAD coronary artery dose compared with the TWB method (55.21 ± 18.68 vs. 60.55 ± 16.05, P = 0.03). Interestingly, the maximum dose (P < 0.0001) and V25 (P < 0.0001) of the heart were both lower in the asymmetric technique rather than the TWB RT. The Dmean of the left lung was also significantly reduced in our institutional RT plan compared with the TWB method (P < 0.0001), while it was comparable between the asymmetric and electron-boosted TWB techniques. The CLD and, consequently, the absolute lung volume under irradiation were the lowest in the asymmetric technique (18 mm) versus TWB and electron-boosted TWB methods.

The average and maximum PTV doses were found to be higher with electron-boosted TWB method compared to the other two RT techniques. No significant difference was found in terms of the mean or maximum dose of PTV between asymmetric and TWB techniques. However, V50 was comparable between all the three methods.

Our findings showed that the asymmetric RT method in our institution improved the DVH parameters of the heart and lung, while it had no significant effects on the PTV coverage.


  Conclusion Top


In the centers where multileaf collimator is not available, the asymmetric technique is useful for patients with left-sided breast cancer, which leads to an acceptable PTV coverage with considerably reduced cardiopulmonary doses. Further studies with longer follow-up period are necessary to identify the long term complications of asymmetric RT technique on heart and lungs.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Fisher B, Anderson S, Bryant J, Margolese RG, Deutsch M, Fisher ER, et al. Twenty-year follow-up of a randomized trial comparing total mastectomy, lumpectomy, and lumpectomy plus irradiation for the treatment of invasive breast cancer. N Engl J Med 2002;347:1233-41.  Back to cited text no. 1
    
2.
Zhang J, Wu S, Sun J, Li F, Guan X, He Z. Comparison of six irradiation techniques for delivering hypofractionated whole-breast radiotherapy with a simultaneous integrated boost after breast-conserving surgery. Int J Radiat Res 2018;16:269-78.  Back to cited text no. 2
    
3.
Wennstig AK, Wadsten C, Garmo H, Fredriksson I, Blomqvist C, Holmberg L, et al. Long-term risk of ischemic heart disease after adjuvant radiotherapy in breast cancer: Results from a large population-based cohort. Breast Cancer Res 2020;22:10.  Back to cited text no. 3
    
4.
Gagliardi G, Bjöhle J, Lax I, Ottolenghi A, Eriksson F, Liedberg A, et al. Radiation pneumonitis after breast cancer irradiation: Analysis of the complication probability using the relative seriality model. Int J Radiat Oncol Biol Phys 2000;46:373-81.  Back to cited text no. 4
    
5.
Vicini FA, Sharpe M, Kestin L, Martinez A, Mitchell CK, Wallace MF, et al. Optimizing breast cancer treatment efficacy with intensity-modulated radiotherapy. Int J Radiat Oncol Biol Phys 2002;54:1336-44.  Back to cited text no. 5
    
6.
Mukesh MB, Barnett GC, Wilkinson JS, Moody AM, Wilson C, Dorling L, et al. Randomized controlled trial of intensity-modulated radiotherapy for early breast cancer: 5-year results confirm superior overall cosmesis. J Clin Oncol 2013;31:4488-95.  Back to cited text no. 6
    
7.
Yavas G, Yavas C, Acar H. Dosimetric comparison of whole breast radiotherapy using field in field and conformal radiotherapy techniques in early stage breast cancer. Int J Radiat Res 2012;10:131.  Back to cited text no. 7
    
8.
Ma CM, Ding M, Li JS, Lee MC, Pawlicki T, Deng J. A comparative dosimetric study on tangential photon beams, intensity-modulated radiation therapy (IMRT) and modulated electron radiotherapy (MERT) for breast cancer treatment. Phys Med Biol 2003;48:909-24.  Back to cited text no. 8
    
9.
Ma CM, Pawlicki T, Lee MC, Jiang SB, Li JS, Deng J, et al. Energy-and intensity-modulated electron beams for radiotherapy. Phys Med Biol 2000;45:2293-311.  Back to cited text no. 9
    
10.
Troy Teo P, Guo K, Ahmed B, Alayoubi N, Kehler K, Fontaine G, et al. Evaluating a potential technique with local optical flow vectors for automatic organ-at-risk (OAR) intrusion detection and avoidance during radiotherapy. Phys Med Biol 2019;64:1-15.  Back to cited text no. 10
    
11.
Yang H, Mir R, Díez P, Tsang Y, Conibear J, Simões R, et al. Provision of organ at risk contouring guidance in uk radiotherapy clinical trials. Clin Oncol (R Coll Radiol) 2020;32:e60-6.  Back to cited text no. 11
    
12.
Tanaka H, Hayashi S, Hoshi H. Determination of the optimal method for the field-in-field technique in breast tangential radiotherapy. J Radiat Res 2014;55:769-73.  Back to cited text no. 12
    
13.
Tanaka H, Ito M, Yamaguchi T, Hachiya K, Yajima T, Kitahara M, et al. High tangent radiation therapy with field-in-field technique for breast cancer. Breast Cancer (Auckl) 2017;11:1-5.  Back to cited text no. 13
    
14.
Hayman JA, Hillner BE, Harris JR, Pierce LJ, Weeks JC. Cost-effectiveness of adding an electron-beam boost to tangential radiation therapy in patients with negative margins after conservative surgery for early-stage breast cancer. J Clin Oncol 2000;18:287-95.  Back to cited text no. 14
    
15.
Mayo C, Lo YC, Fitzgerald TJ, Urie M. Forward-planned, multiple-segment, tangential fields with concomitant boost in the treatment of breast cancer. Med Dosim 2004;29:265-70.  Back to cited text no. 15
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]
 
 
    Tables

  [Table 1], [Table 2]



 

Top
   
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
   Abstract
  Introduction
   Materials and Me...
  Results
  Discussion
  Conclusion
   References
   Article Figures
   Article Tables

 Article Access Statistics
    Viewed130    
    Printed2    
    Emailed0    
    PDF Downloaded21    
    Comments [Add]    

Recommend this journal